When you measure nonlinearity, spurious responses, or SFDR, your first job is to ensure the source is not producing artifacts that you mistakenly attribute to the DUT. In microwave testing, this is a common failure mode: a tone appears on the analyzer and the team starts debugging the DUT—when the source spectrum or the setup is the real cause.
For a wideband reference platform, see the CAMS9K40G analog RF signal generator.
Three artifact families engineers watch
- Harmonics: multiples of the carrier (2f, 3f…).
- Sub-harmonics: fractional responses tied to synthesis paths and internal architecture.
- Non-harmonic spurs: discrete tones not at harmonic positions, often the easiest to misinterpret.
Common causes of false responses
- Insufficient filtering between source and DUT.
- Mismatch/reflections causing ripple or mixing in the chain.
- Analyzer overload creating its own intermod products.
- Cable routing issues (trigger/digital lines coupling into RF).
Mitigation strategies that work
- Validate source cleanliness at the DUT plane, not just at the generator output.
- Use bandpass filtering near the DUT when testing wideband devices or high dynamic range scenarios.
- Control analyzer input level with appropriate attenuation to avoid overload.
- Keep trigger and control lines physically separated from RF paths.